चिकित्सा उपकरण मरम्मत मामले स्ट्राइकर 30o 0502-503-030 मरम्मत के लिए लैप्रोस्कोप
Stryker 30º 0502-503-030 laproscope, SN:103**** Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. Repair cost: $*** Common Issues with the Stryker ...
STORZ 1HD H3-ZA 22220061 ऑटोक्लेवेबल कैमरा हेड TH104 केबल
Storz 1HD H3-ZA 22220061 / TH104 Autoclavable Camera Cable Brand: Storz Model: H3-ZA 22220061 Series: Camera Cable Use for Karl Storz Image 1 Spies HD Camera Head H3-ZA TH104 TH-104 If you have any questions about 1HD H3-ZA Autoclavable Camera Cable 22220061 for Storz, please feel free to contact us...
चिकित्सा उपकरण की मरम्मत के लिए जीई आरआईसी 5-9-डी अल्ट्रासाउंड जांच के मामले
GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in ...
OLYMPUS CF-HQ190L चिकित्सा उपकरण मरम्मत मामले उपकरण मरम्मत मामले
OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not ...
मरम्मत के लिए PENTAX मेडिकल EPK-i7000 वीडियो प्रोसेसर
PENTAX Medical EPK-i7000 Video Processor For Repair ► Brief Description Brand: PENTAX (PENTAX Endoscope Processor) Model: EPK-i7000 Serial Number: AC*** Fault: It cannot recognize the endoscope and reports an error. Solution: Repair the system board. Repair Time: 3-5 working days. Warranty Period: ...
मरम्मत के लिए ओलिंपस एलटीएफ-160 थोरकोस्कोप
Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the ...
मरम्मत के लिए OLYMPUS CF-H180AL इलेक्ट्रॉनिक कोलोनोस्कोप
OLYMPUS CF-H180AL Electronic Colonoscope For Repair ► Brief Description Inspection Results: 1. The image is dark, and the CCD is aged. 2. One branch of the light beam is disconnected. 3. The insertion tube is aged and wrinkled. 4. The lower angle is malfunctioning, the other angles are tight, and ...
मरम्मत के लिए जीई 4डी3सी-एल 4डी उत्तल ट्रांसड्यूसर
GE 4D3C-L 4D Convex Transducer For Repair ► Brief Description Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Solution:Replace the dome, dome tube; repair electric motor. ►Problem & Solutoin BrandGEModel4D3C-LFault1.Probe dome ...
मरम्मत के लिए GE 4C-RS पेट का जांच
GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS ...